The test results of the system verify the sampling resistance circuit. Refer to "JJG L66 93 dc resistor" verification regulation, volt ampere method is used to calibrate it to verify whether the accuracy of the multi-function calibration system meets the design specifications. The adjustable constant current source provides the test current, the range is 0~600 A, and the current converter and voltmeter together make up the current measurement standard. The composition of the whole calibration scheme test system
The sampling resistance error is the actual sampling resistance value under different test current at 75 m Omega, and its relative error is less than 0.003%. Table 4 shows the error of simulating different resistance values of high power resistance when the fixed test current is 100 A. It can be seen from the table that the relative error is within 0.004%. This is because the device system uses such as DC comparator, differential amplifier, sampling signal amplifier, D/A converter is a high linearity element, high linearity system. It is proved that the accuracy of the system analog output resistance can be within 0.02% range and meets the design requirements after several tests and tests under different values and current.
表3 75mΩ在不同输入电流时的误差 | ||||||||
实际电流/A | 29.9996 | 40.0001 | 50.0001 | 59.9999 | 70.0008 | 80.0006 | 89.9995 | 100.0019 |
实际电阻/mΩ | 74.9996 | 74.9977 | 75.0002 | 75.0002 | 74.9997 | 74.9992 | 75.0016 | 75.0015 |
误差/% | -0.0005 | -0.003 | 0.0003 | 0.0003 | -0.0004 | -0.0010 | 0.0021 | 0.0019 |
表4 100A测试电流下不同电阻值的误差 | |||||||
模拟电阻/mΩ | 7.5 | 22.5 | 30.0 | 45.0 | 60.0 | 67.5 | 75.0 |
实际电阻/mΩ | 7.5003 | 22.5001 | 30.0011 | 45.0009 | 60.0006 | 67.5001 | 75.0015 |
误差/% | 0.004 | 0.0004 | 0.0037 | 0.002 | 0.001 | 0.0001 | 0.0019 |
The block diagram of the calibration device for the sampling resistance multifunction calibration system. Data analysis, multi-functional calibration system calibration device for the system to calibrate the measurement. Current measurement error: Table 1 is relative error at different current tests under 20 A range, and the relative error is less than 0.011%. Relative error of the relative error at different current tests for 100 A range is less than 0.01%。
表1 20A量程下在不同电流测试时的误差 | |||||||||
实际电流/A | 4.0000 | 6.0000 | 8.00009 | 10.00002 | 12.0000 | 14.00004 | 15.99999 | 18.00000 | 20.00015 |
测量电流/A | 3.9996 | 5.9996 | 7.9996 | 9.9996 | 11.9996 | 13.9996 | 15.9997 | 17.9997 | 20.0000 |
误差 /% | -0.0102 | -0.0067 | -0.0062 | -0.0042 | -0.0034 | -0.0031 | -0.0018 | -0.0017 | -0.0007 |
表2 100A量程下在不同电流测试时的误差 | ||||||||
实际电流/A | 29.9996 | 40.0001 | 50.0001 | 59.9999 | 70.0008 | 80.0006 | 89.9995 | 100.0019 |
测量电流/A | 29.998 | 39.998 | 49.999 | 59.999 | 69.999 | 79.998 | 89.999 | 100.001 |
误差/% | -0.0054 | -0.0052 | -0.0022 | -0.0014 | -0.0026 | -0.0033 | -0.0006 | -0.0009 |
Conclusion the design of multifunctional sampling calibration system for resistance tester simulation resistor in the traditional current mode active innovation, using precision simulation technology and high resistance current measurement technology, and three types of instrument verification device with a wide range of resistance value can be set up, which can simulate the continuous 1U ~3.75 arbitrary resistance value, the resolution of 10N. And 0.02 grade accuracy; test current range, up to 600 A and 0.02 level accuracy, related technical indicators have reached the domestic leading level in the industry, to meet the requirements of three kinds of test instruments, verification has been widely used on resistance tester in the loop resistance tester, DC resistance tester and grounding to guide the。